Skip to content

Public FDA record

Z-2603-2018: Medtronic CD HORIZON Spinal System is intended to help provide immobilization and stabilization of spinal

Recall record Z-2603-2018 for Medtronic Sofamor Danek USA Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2603-2018
Company in FDA record
Medtronic Sofamor Danek USA Inc
FDA record creation date
Aug 3, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

This lot has incorrectly been laser marked both on the screw head itself as well as the data carrier tag attached to the screw. The description on face two of the screw head and on the data carrier tag both incorrectly indicate a 5.5 mm diameter screw when the actual screw diameter aligns with 5.0 diameter screw size requirements.

Source facts

Action
Medtronic distributed a voluntary removal notice via FedEx courier service to impacted U.S. consignees, delivered 09 July 2018. The notice stated provided a list of Loaner kits where the screws may also be found and provided the following instructions: " Instructions to Customers: 1.) Please locate and remove the impacted product from normal storage locations. Do not use this product. 2.) Complete the enclosed Customer Confirmation Form and return via email to rs.sdmriskmgt@medtronic.com. 3.) As applicable, contact your Medtronic Sales Representative to coordinate the return and replacement of affected product. Transmission of this Field Safety Notice: 1.) This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. 2.) Please transfer this notice to other organizations on which this action has an impact. 3.) Please maintain a copy of this notice in your records."
Initiated
2018-07-09
Root cause
Employee error
Terminated
2021-05-07
Distribution
US
Recall number
Z-2603-2018
Recall status
Terminated
CDRH record ID
165709
FDA record created
2018-08-03
Recall event number
80541

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗