This is a retrospective recall of an instrument used in spine fusion procedures with Modera Modular Pedicle Screw system implants. The firm received a customer complaint in which excessive forces were applied to the shank insertion instrument and the laser weld failed.
Source facts
Action
On February 13, 2017, the firm emailed its distributors to notify them that their Rev. A devices would be replaced with Rev. B devices, and asked the distributors to mail back the Rev. A devices.
Initiated
2017-02-13
Root cause
Device Design
Terminated
2018-11-07
Distribution
2 distributors in CO and TX.
Recall number
Z-2604-2018
Recall status
Terminated
CDRH record ID
165627
FDA record created
2018-08-03
Recall event number
80499