Z-2605-2011: Access Immunoassay Systems Ostase Calibrators, Part Number: 37305 The Access Ostase Calibrators are used

Recall record Z-2605-2011 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2605-2011
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Jun 21, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The recall was initiated because Beckman Coulter has confirmed that Access Ostase Calibrator and QC vials from the affected lots have shown an increased rate of fracturing and may leak the contents in the box during shipment to customers. Operator injury may occur due to contact with broken glass, exposure to infectious agents, and/or exposure by skin contact to agents that may cause sensitizatio

Source facts

Action
Beckman Coulter sent an PRODUCT CORRECTIVE ACTION (PCA) letter on May 10, 2011 to all affected customers who purchased Access Ostase Calibrators and Access Ostase QC. The letter identified the customers with an explanation of the problem and the potential impact should the problem occur and the action needed to be taken. Customer were instructed to: (1) Inspect all Access Ostase Calibrator and QC vials and packaging for signs of failure prior to use, including: -Cracking, particularly along the bottom of the vial, -Absence of reagent and/or leaking of contents into the box (2) Human source material used in the preparation of Access Ostase Calibrator and QC material has been tested and found non-reactive for Hepatitis B (HBV), Hepatitis C (HCV), and Human Immunodeficiency Virus (HIV-1 and HIV-2). No known test method can offer complete assurance that infectious agents are absent. Access Ostase Calibrator and QC vials should be handled in accordance with Good Laboratory Practices, and should be handled as if capable of transmitting infectious disease. (3) Skin contact with Access Ostase Calibrator and QC material may cause sensitization. Wear suitable gloves when handling vials. If contact with skin occurs, wash immediately with plenty of soap and water. Customers were told to share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected products listed above to another laboratory, please provide a copy of the letter to them. In addition, customers were requested to complete and return the enclosed Response Form within 10 days so we are assured you have received this important communication. For replacement of defective vials: (1) In the United States, please contact Client Services at 1-800-526-3821, option 1. (2) In Canada, please contact Customer Service at 1-800-463-7828. (3) Outside of the United States and Canada, contac
Initiated
2011-05-10
Root cause
Package design/selection
Terminated
2012-07-11
Distribution
Nationwide in US and worldwide to: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Italy, Japan, Jordan, Lebanon, Liechtenstein, Mexico, Morocco, New Zealand, Panama, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom
Recall number
Z-2605-2011
Recall status
Terminated
CDRH record ID
100389
FDA classified
2011-06-21
Recall event number
58833

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗