Skip to content

Public FDA record

Z-2605-2018: Think Surgical Cutter, Flat, 6.2mm x 144mm. For use only with the TCAT Computer-Assisted Tool. Part Num

Recall record Z-2605-2018 for Think Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2605-2018
Company in FDA record
Think Surgical, Inc.
FDA record creation date
Aug 6, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for the cutter head to detach from the cutter shaft.

Source facts

Action
The firm, Think Surgical Inc., sent an "Urgent Medical Device Recall" notices dated 3/30/2018 to customers on 3/30/18. The notices instructed customers to perform the following actions: Examine your inventory, locate any unused devices noted above and quarantine them immediately. If you have received any accessory kits from THINK surgical Incorporated with the noted cutter part numbers included, do not use any affected product and return any unused product to the company. THINK Surgical Incorporated has provided you with a Returned Material Authorization Form (RMA), a FedEx Envelope and return FedEx AirBill to retun any unused devices. Your account will receive a credit and replacement units will be sent. Fill out the accompanying Acknowledgement Form. Email the completed Acknowledgement Form to Ms. Meliha Mulalic at mmulalic@thinksurgical.com. Contact THINK Surgical Incorporated QA/RA Group at 510-249-2337, attention Ms. Meliha Mulalic, should you have any questions. Urgent Medical Device Recall Information: Lot Number Correction letters dated 4/4/18 were distributed as well. The letters corrected an error in the initial notification letter which alerted customers to a typographical error in which lot 9159 should read 9519.
Initiated
2018-03-30
Root cause
Process control
Terminated
2020-01-24
Distribution
The devices were distributed to the following US states: NJ, NY, OH, and TX. The devices were also distributed to Korea.
Recall number
Z-2605-2018
Recall status
Terminated
CDRH record ID
164544
FDA record created
2018-08-06
Recall event number
80091

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗