Open original FDA record ↗Recall record
- FDA identifier
- Z-2605-2018
- Company in FDA record
- Think Surgical, Inc.
- FDA record creation date
- Aug 6, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential for the cutter head to detach from the cutter shaft.
Source facts
- Action
- The firm, Think Surgical Inc., sent an "Urgent Medical Device Recall" notices dated 3/30/2018 to customers on 3/30/18. The notices instructed customers to perform the following actions: Examine your inventory, locate any unused devices noted above and quarantine them immediately. If you have received any accessory kits from THINK surgical Incorporated with the noted cutter part numbers included, do not use any affected product and return any unused product to the company. THINK Surgical Incorporated has provided you with a Returned Material Authorization Form (RMA), a FedEx Envelope and return FedEx AirBill to retun any unused devices. Your account will receive a credit and replacement units will be sent. Fill out the accompanying Acknowledgement Form. Email the completed Acknowledgement Form to Ms. Meliha Mulalic at mmulalic@thinksurgical.com. Contact THINK Surgical Incorporated QA/RA Group at 510-249-2337, attention Ms. Meliha Mulalic, should you have any questions. Urgent Medical Device Recall Information: Lot Number Correction letters dated 4/4/18 were distributed as well. The letters corrected an error in the initial notification letter which alerted customers to a typographical error in which lot 9159 should read 9519.
- Initiated
- 2018-03-30
- Root cause
- Process control
- Terminated
- 2020-01-24
- Distribution
- The devices were distributed to the following US states: NJ, NY, OH, and TX. The devices were also distributed to Korea.
- Recall number
- Z-2605-2018
- Recall status
- Terminated
- CDRH record ID
- 164544
- FDA record created
- 2018-08-06
- Recall event number
- 80091