Z-2605-2026: Omnipod (Eros) Pod. Includes the below Model/REF Numbers: 1. POD-ZXP425 (10 Pack), POD- ZXP420 (Single

Recall record Z-2605-2026 for Insulet Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2605-2026
Company in FDA record
Insulet Corporation
FDA classification date
Jun 30, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

Source facts

Action
Insulet initially issued a press release on 05/26/2026 along with an "Urgent: Medical Device Correction" on the same day. Consignees were notified of the issue and hazard and instructed to not used Pods from affected lots and to check the lot number on the Pod tray lid, box or Pod itself to determine if affected Pods are on hand. If impacted Pods are on hand, consignees were instructed to discontinue use of any impacted Pod immediately, immediately change the Pod if currently using an affected lot, and acknowledge the email confirming receipt. Confirming receipt of the email is needed even if no affected Pods are on hand. Customers will receive a return kit with clear instructions on how to send back any unused affected Pods.
Initiated
2026-05-20
Root cause
Process design
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey.
Recall number
Z-2605-2026
Recall status
Open, Classified
CDRH record ID
220616
FDA classified
2026-06-30
Recall event number
99034

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗