Z-2609-2025: FilmArray NGDS Warrior Panel NGDS-ASY-0007 is a qualitative, multiplexed, nucleic acid-based in vitro dia

Recall record Z-2609-2025 for Biofire Defense. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2609-2025
Company in FDA record
Biofire Defense
FDA classification date
Sep 19, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

FilmArray NGDS Warrior Panel (NGDS-ASY-0007) has an increased risk of internal control failures when testing positive blood culture, which may lead to delayed results and additional diagnostic workup.

Source facts

Action
On 7/10/2025, correction notices were mailed and emailed to customers who were asked to do the following: - If testing with only whole blood and/or sputum, there is no impact to the product s performance. Continue using panel kits as planned. - If testing with positive blood culture may occur, discontinue use, dispose of impacted kit(s) according to your lab s procedures, and only use replacement kits provided, or use impacted lots only with whole blood and/or sputum sample types. - Distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties who may use this product, including others to whom you may have transferred our product. - Complete and return the acknowledgment of receipt form via email to support@biofiredefense.com
Initiated
2025-07-10
Root cause
Component design/selection
Distribution
US Nationwide distribution in the states of AL, AZ, UT, CA, TX, DC , VA, FL, MD, OH, HI, NY, KY, LA, MA, MI, MN, MO, NC, NH, RI, NV, TN, IA, WA.
Recall number
Z-2609-2025
Recall status
Open, Classified
CDRH record ID
215069
FDA classified
2025-09-19
Recall event number
97286

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗