Skip to content

Public FDA record

Z-2616-2014: Streamline MIS Spinal Fixation System Fixed Rod Holder NON STERILE, Rx only The Streamline MIS Spinal

Recall record Z-2616-2014 for PIONEER SURGICAL TECHNOLOGY, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2616-2014
Company in FDA record
PIONEER SURGICAL TECHNOLOGY, INC.
FDA classification date
Sep 9, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The affected fixed rod holders did not allow the instrument to lock down on the spinal rod.

Source facts

Action
Pioneer Surgical sent a Product Field Action letter on July 7, 2013, to all affected customers. The letter described the affected product and issue, and provided instructions for return of the recalled product. All 9 affected instruments were returned to Pioneer Surgical, reconditioned and sent back to the field. Customers with questions were instructed to call 906-226-4489. For questions regarding this recall call 906-226-4812.
Initiated
2013-06-14
Root cause
Device Design
Terminated
2014-09-30
Distribution
Nationwide Distribution including TX, CA, IL, and FL.
Recall number
Z-2616-2014
Recall status
Terminated
CDRH record ID
129377
FDA classified
2014-09-09
Recall event number
69036

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗