Z-2616-2021: Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-

Recall record Z-2616-2021 for Datascope Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2616-2021
Company in FDA record
Datascope Corp.
FDA record creation date
Sep 30, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs during a very specific set of conditions.

Source facts

Action
On Wednesday, September 8, 2021, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that all lots of Maquet Cardiosave Hybrid experience the following issue: Unexpected Shutdown Upon Battery Removal. Immediate Actions to be taken by Customers: To ensure that the Cardiosave Hybrid is not susceptible to shutdown during the removal of a battery, users should not eject a battery when: - There is a singulat battery in either battery charging bay and unused battery bay is empty. - The battery is charging on AC power (indicated by a flashing LED on the battery pack). Additionally, the user can prevent the issue by ensuring that a battery is inserted into each of the two battery bays. Should a charging battery be ejected inadvertently, the unit will automatically switch over to the second battery, thus preventing system shutdown. Furthermore, Datascope/Getinge is developing a software correction to address this issue. A Datascope/Getinge service representative will contact you to schedule the installation of the updated software. This work will be done at no cost to your facility. If you have any questions, please contact your Datascope/Getinge representative or call the Datascope/Getinge Technical Support at 1-888-943-8872, options 4, 2, 1, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Initiated
2021-09-08
Root cause
Software design
Distribution
Nationwide domestic distribution. Foreign distribution worldwide.
Recall number
Z-2616-2021
Recall status
Open, Classified
CDRH record ID
189204
FDA record created
2021-09-30
Recall event number
88664

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗