Open original FDA record ↗Recall record
- FDA identifier
- Z-2618-2025
- Company in FDA record
- Boston Scientific Corporation
- FDA classification date
- Sep 23, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The product in incorrectly labeled. The label indicates that the skive hole should be above the balloon, while it is actually positioned below, and vice versa.
Source facts
- Action
- On August 28, 2025 URGENT MEDICAL DEVICE REMOVAL letters were sent to customers. Instructions: 1. Immediately cease further distribution or use of any remaining affected devices. Segregate the units in a secure place until they can be returned. 2. Immediately post this information in a visible location near the affected devices to ensure it is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization and to any facilities where affected devices have been transferred, including hospitals or sites within your network. 4. Return affected devices.
- Initiated
- 2025-08-28
- Root cause
- Process control
- Distribution
- US Nationwide distribution in the states of CA, CO, GA, ID, IL, IN, MA, ME, MI, MS, NC, OH, OK, PA, SD, TN, TX, VA, WA.
- Recall number
- Z-2618-2025
- Recall status
- Open, Classified
- CDRH record ID
- 215852
- FDA classified
- 2025-09-23
- Recall event number
- 97529