Open original FDA record ↗Recall record
- FDA identifier
- Z-2618-2026
- Company in FDA record
- Medacta Usa Inc
- FDA classification date
- Jun 29, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.
Source facts
- Action
- On 4/30/2026, field safety notices were emailed to customers who were asked to do the following: No pre-emptive interventions are recommended, however monitoring of the range of motion and stability should be evaluated during patient follow-up visits as per the standard of care. " Return all units of the lot numbers listed to firm. " Forward this notice to all people and organizations potentially affected. Firm Customer Service can be reached at 312-878-2381.
- Initiated
- 2026-04-30
- Root cause
- Device Design
- Distribution
- US Nationwide distribution in the states of CO, AR, IN, PA, UT, FL.
- Recall number
- Z-2618-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220256
- FDA classified
- 2026-06-29
- Recall event number
- 98944