Open original FDA record ↗Recall record
- FDA identifier
- Z-2619-2024
- Company in FDA record
- DKK Dai-Ichi Shomei Co., Ltd.
- FDA classification date
- Aug 14, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The weld on the light head frame can crack which could lead to problems such as the light head frame sagging, the plastic front face housing cracking, or the light head detaching.
Source facts
- Action
- On July 9, 2024 URGENT: Aurora 7 Lighthead Series RECALL letters were sent to customers. Actions to be taken by the customer: Contact the end users and tell them to discontinue the use of the product. Please send an acknowledgement letter or email to Dai-ichi Shomei to let us know that you have received this Field Action Advisory Notice within 3 days after you receive the Notice. Actions to be Taken by the Manufacturer: The physical correction will be announced later. Contact information for questions: email info@hospilite.co.jp
- Initiated
- 2024-07-09
- Root cause
- Process control
- Distribution
- Domestic: Nationwide Distribution International: Australia, Canada, Egypt, Hong Kong, Indonesia, Japan, Jordan, Netherlands, New Zealand, Philippines, Singapore, South America, South Korea, Thailand, UAE.
- Recall number
- Z-2619-2024
- Recall status
- Open, Classified
- CDRH record ID
- 209004
- FDA classified
- 2024-08-14
- Recall event number
- 94979