Open original FDA record ↗Recall record
- FDA identifier
- Z-2619-2026
- Company in FDA record
- Windstone Medical Packaging, Inc.
- FDA classification date
- Jul 1, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Source facts
- Action
- AMS ALIGNED MEDICAL SOLUTIONS issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 5/21/2026 via email. The notice explained the issue, potential risk, and requested the following: "AMS requests that you undertake the following activities: 1. Immediately check your inventory for the recalled products listed above and place them under quarantine. 2. Due to the manufacturing process any finished sterilized packs remaining in inventory, or at your facility will require over labeling for all products affected by this recall. Packs will be labeled with a sticker attached to each pack. This sticker will read: RECALL NOTICE CARDINAL HEALTH announced the recall of 5110 Webcol" Large Alcohol Prep Pad contained in this kit after it was sealed and sterilized. All other components in the kit are not affected by this recall. " At the time the kit is opened for use any Cardinal Alcohol Prep Pad should be identified and set aside. " The recalled Cardinal Alcohol Prep Pad should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution. 3. Complete the attached Recall Reply Form listing all inventory of affected product by lot number and quantity. 4. If you have further distributed this product, please identify your distribution partners, and notify them at once of the product recall. Your notification to your customers should include a copy of this notification letter. 5. Please complete the Recall Reply Form within 5 business days and return via email to fieldcorrectiveaction@alignedmedical.com even if you have no affected product on hand." For questions or concerns: fieldcorrectiveaction@alignedmedicalsolutions.com.
- Initiated
- 2026-05-21
- Root cause
- Material/Component Contamination
- Distribution
- US Nationwide distribution in the states of PA, GA, CA, WA.
- Recall number
- Z-2619-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221001
- FDA classified
- 2026-07-01
- Recall event number
- 99189