Z-2622-2025: BeneVision N1 Patient Monitor (N1). Part Numbers: 6660E-PA00017, 6660E-PA00014, 6660E-PA00012, 6660E-PA00

Recall record Z-2622-2025 for Mindray DS USA, Inc. dba Mindray North America. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2622-2025
Company in FDA record
Mindray DS USA, Inc. dba Mindray North America
FDA classification date
Sep 25, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for activation of an abnormal alarm pause.

Source facts

Action
Mindray contacted consignees via letter on about 08/15/2025. consignees were instructed that they may continue to use the BeneVision N1 patient monitor as a multi-parameter module with Mindray host monitors. Per the additional guidance, it is recommended to confirm that the remaining alarm time counter is actively displayed and follows the existing N1 Operator's Manual: Instructions for when the Alarm Pause is Activated. An image was also included in the notification letter showing the correctly displayed Alarm Pause notification and timer. Consignees were also requested to make all applicable clinical staff aware of the issue and to complete and return the Acknowledgement and Receipt form. Mindray's Care team will deploy a software update to correct the issue. Distributors were advised to contact their direct customers and notify them directly.
Initiated
2025-08-15
Root cause
Software design
Distribution
Worldwide - Nationwide distribution and the country of Canada.
Recall number
Z-2622-2025
Recall status
Open, Classified
CDRH record ID
215600
FDA classified
2025-09-25
Recall event number
97437

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗