Z-2624-2026: Medtronic Harmony Delivery Catheter System

Recall record Z-2624-2026 for Medtronic Heart Valves Division. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2624-2026
Company in FDA record
Medtronic Heart Valves Division
FDA classification date
Jul 2, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a manufacturing issue.

Source facts

Action
Medtronic issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 5/28/2026 via mail or email. The notice explained the issue, potential risk, and requested the following: " - Review your inventory for affected lot numbers listed. - If unused inventory from the listed lot numbers is located, immediately quarantine and return the affected units to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return and exchange or credit of the affected units. Your Medtronic sales representative can assist in the return of the affected units as necessary. - Please share this notification with implanters and teams within your organization. If product listed above has been forwarded from your facility to another facility, please notify the other facility of this Medtronic Urgent Medical Device Recall. - Complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com. This form must be returned even if you do not have the affected units in your possession. - Please maintain a copy of this communication in your records."
Initiated
2026-05-22
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Canada, Chile, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Mexico, Netherlands, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Recall number
Z-2624-2026
Recall status
Open, Classified
CDRH record ID
220767
FDA classified
2026-07-02
Recall event number
99031

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗