Open original FDA record ↗Recall record
- FDA identifier
- Z-2625-2016
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Aug 23, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to an internal communication error between the firmware and the software of the components, the planned CT scan executes properly, but the injector is not started. Therefore, the contrast agent is not injected and the desired examination result is not achieved. This error only affects the automatic mode or coupled mode and does not affect the manual control of the injector.
Source facts
- Action
- Siemens mailed to customers an Important Customer Safety Notice to inform about a possible malfunction when using the affected systems and what actions to take to eliminate the problem. Also, Siemens informed their customers that they were currently developing a new firmware to eliminate the internal communication error, which should be released in August 2016. Customers were asked to contact their service organization at 800-888-7436, if they had any questions.
- Initiated
- 2016-07-13
- Root cause
- Software in the Use Environment
- Terminated
- 2017-01-24
- Distribution
- Distributed to: MI, NY, CA, KY, ND, NE
- Recall number
- Z-2625-2016
- Recall status
- Terminated
- CDRH record ID
- 148569
- FDA classified
- 2016-08-23
- Recall event number
- 74840