Z-2625-2023: Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007

Recall record Z-2625-2023 for Remote Diagnostic Technologies Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2625-2023
Company in FDA record
Remote Diagnostic Technologies Ltd.
FDA classification date
Sep 21, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Full screen error message may occur on patient monitors with affected hardware version, either during or after laryngoscopes are unplugged from monitors, informing users a 60-100 second shutdown/restart are required. Error prevents: viewing graphical patient vital signs, sounding of pulse audio tone, visualization of patient airway used during intubation; may lead to treatment delay or hypoxia.

Source facts

Action
On 8/24/23 Correction Notices were mailed to customers and distributors who were asked to do the following: 1) If monitor has Trizeps-7 Hardware, remove the laryngoscope from service with this monitor. Users must use an alternative laryngoscope not connected to the Tempus Pro Monitor to manage the patient s airway to avoid interruption in patient care. The Tempus Pro Monitor can remain in service if the Tempus Pro Video Laryngoscope is not connected to the monitor. 2) If monitor has Trizeps-6 Hardware, the laryngoscope can continue to be used with the monitor. 3) Post this Urgent Medical Device Correction letter on or near the affected device. 4) Pass this notice on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred (if appropriate). 5) Complete and return the Urgent Medical Device Correction response form. Distributors were asked to take the following additional actions: If you have any Tempus Pro Video Laryngoscopes or any Tempus Pro Monitors in stock that have not been distributed, do not distribute them. Modify the response form to include your firm's contact information and then send the response form along with the customer correction notice to customers. If you need any further information or support contact the firm at (800) 722-9377 or email RDT.RecalI.Response@Philips.com. On 11/6/23, updated correction notices were distributed adding two additional models that could become affected if their hardware is updated. Additionally, the notices asked customers to install software (V07.34) on Tempus Pro Monitors according to installation instructions.
Initiated
2023-08-24
Root cause
Device Design
Distribution
US:AZ, AL, CA, CO, DC, DE, FL , GA, HI, IA, ID, IL, IN , KS, KY, MA , MD , ME, MI, MN, MO, MS, MT, NC , ND, NE, NJ, NM , NV, NY, OH, OK, OR, PA, TN, TX, UT, VA , WA, WI, WV, WY. OUS: AE, AT, AU, BE, BN, BO, CA, CH, CZ, DE, DK, ES, FR, GB, HK, IE, IL, IS, IT, KY, LU, MY, NL, NO, OM, PT, RO, SA, SE, SG, SI, TH, YT, ZA
Recall number
Z-2625-2023
Recall status
Open, Classified
CDRH record ID
202893
FDA classified
2023-09-21
Recall event number
92914

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗