- FDA identifier
- Z-2625-2026
- Company in FDA record
- Hamilton Medical AG
- FDA classification date
- Jul 1, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.
Source facts
- Action
- On May 22, 2026, Hamilton Medical issued a Urgent Medical Device Recall to affected consignees. Hamilton Medical ask consignees to take the following actions: 1. Identify affected devices (consider devices in use, in stock, spare devices etc.) and make them available for exchange of cuff connectors. 2. Continue to use IntelliCuff devices according to their labelling. Mitigate potential risks by following the actions described in point B. Type of Action to mitigate the risk . 3. Please fill in and sign Customer Reply Form (page 5) and send it to your Hamilton Medical distribution partner or subsidiary. This must be conducted as quickly as possible, but no later than 30 calendar days after receiving the Field Safety Notice.
- Initiated
- 2026-05-22
- Root cause
- Process design
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IL, KY, MA, MI, ND, NE, NE, NY, OH, PA, PR, SD, TX, VA, WA, WI, WV, WY and the country of Canada.
- Recall number
- Z-2625-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220706
- FDA classified
- 2026-07-01
- Recall event number
- 99035