Open original FDA record ↗Recall record
- FDA identifier
- Z-2626-2017
- Company in FDA record
- EOS Imaging
- FDA record creation date
- Jun 23, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
ROTH MOBELI grab handles could detach and patient could fall if the EOS system operator does not follow instructions. Before each use, the operator must examine the vacuum safety indicator to check the vacuum is monitored and sufficient and the grab handle correctly attached. Do not use grab handles to help patient up. The handles are only for stabilizing the patients hand or arm in the EOS cabin.
Source facts
- Action
- EOS Imaging sent Field Safety Notice letter and Removal Instructions May 30, 2017. Customers are to immediately discontinue use of the grab handles and remove them from all affected units from inventory and return to a EOS Imaging. For further questions please call (678) 564-5400.
- Initiated
- 2017-05-30
- Root cause
- Error in labeling
- Terminated
- 2018-10-09
- Distribution
- US Distribution to the states of : Pennsylvania, Massachusetts, Minnesota, Alabama, Michigan, Texas and Washington.
- Recall number
- Z-2626-2017
- Recall status
- Terminated
- CDRH record ID
- 156038
- FDA record created
- 2017-06-23
- Recall event number
- 77424