Z-2626-2023: FLUIDSHIELD 3 N95 Particulate Filter Respirator and Surgical Mask with SO SOFT* Lining, Orange, size smal

Recall record Z-2626-2023 for O&M HALYARD, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2626-2023
Company in FDA record
O&M HALYARD, INC.
FDA classification date
Sep 21, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

One respirator sample failed a filtration efficiency test. This respirator may not provide adequate protection to the user.

Source facts

Action
An "URGENT: PRODUCT FIELD ACTION" dated 8/9/23 was sent to customers. The letter described the product, the problem and actions to take. The customers were instructed to do the following: If you received the lot identified above, please review your inventory, and segregate and quarantine all affected product. All other codes and lots of this product are NOT impacted by this Field Action. Customers (end-users): 1. REVIEW your inventory for the affected product code and specific lot number. 2. SEGREGATE AND QUARANTINE all product on-hand per the Affected Product table above. 3. DESTROY or DISCARD the product per the facility s standard disposal procedure for face masks. 4. COMMUNICATE this Field Action to staff to ensure awareness of it. 5. RETURN the completed Field Action Response Form and Certificate of Destruction Form to OMRA_RECALLS@owens-minor.com as soon as possible even if you have no current product inventory. 6. Any adverse events associated with the use of the affected product should be reported to PIQ@hyh.com. 7. If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. If you have any questions, please email Anu Singh @ OMRA_RECALLS@owens-minor.com using the Event # FA-2023-011 in the header of the response. This will ensure prompt response by the O&M Halyard, Inc. team.
Initiated
2023-08-09
Root cause
Under Investigation by firm
Distribution
US Distribution to States of: FL, GA, IA, IL, MO, MS, NE, OK, TN, TX, WI and OUS International to country of: Columbia
Recall number
Z-2626-2023
Recall status
Open, Classified
CDRH record ID
202959
FDA classified
2023-09-21
Recall event number
92942

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗