Open original FDA record ↗Recall record
- FDA identifier
- Z-2627-2011
- Company in FDA record
- Becton Dickinson & Co.
- FDA classification date
- Jun 23, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
In vitro diagnostic instrumentation may exhibit random operational errors that could delay effective patient treatment.
Source facts
- Action
- BD Diagnostics Systems sent an "URGENT PRODUCT RECALL" letter dated August 2010 to all affected customers. The letter describes the product, problem and action to be taken. The letter described the type of error messages the customers may experience. All affected units will be scheduled for on site software update as well as a disk drive replacement. A BD field service engineer will contact the customers to schedule the on site visit. Customers experiencing issues prior to the scheduled visit or have questions should contact the BD Diagnostics Service Department at 800-638-8663.
- Initiated
- 2010-08-10
- Root cause
- Software design
- Terminated
- 2011-07-05
- Distribution
- Worldwide Distribution--USA (nationwide) including countries of Australia, Belgium, Canada, China, Colombia, New Zealand, Hong Kong, Japan, Korea, Philippines, Singapore, Taiwan, and Thailand.
- Recall number
- Z-2627-2011
- Recall status
- Terminated
- CDRH record ID
- 100454
- FDA classified
- 2011-06-23
- Recall event number
- 58873