Open original FDA record ↗Recall record
- FDA identifier
- Z-2627-2016
- Company in FDA record
- Merge Healthcare, Inc.
- FDA classification date
- Aug 23, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for RadSuite AV viewer to skip image slices when the user presses page up or page down when the viewer displays multiple viewports and the user has mouse focused on any viewport other than the first viewport.
Source facts
- Action
- Merge Healthcare sent an Urgent Medical Device Recall letter dated January 28, 2016, and via email on January 31, 2013, to all affected customers. the letter identified the product the problem and the action needed to be taken by the customer. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. In addition, the customer is requested to call Customer Service or send them an email to schedule the upgrade. If the customer chooses not to upgrade, they are requested to complete the attached form or go online and complete it. If you have any additional questions call Merge Customer Service at (877) 741-5369. For further questions, please call (262) 367-0700.
- Initiated
- 2016-01-31
- Root cause
- Software design
- Terminated
- 2017-02-10
- Distribution
- Distribution was made to AL, AZ, CA, IL, MD, MA, MI, MO, NV, NJ, NC, OH, PA, TN, and VA. Government distribution was made to OK. There was no foreign/military distribution.
- Recall number
- Z-2627-2016
- Recall status
- Terminated
- CDRH record ID
- 148776
- FDA classified
- 2016-08-23
- Recall event number
- 74919