Open original FDA record ↗Recall record
- FDA identifier
- Z-2628-2016
- Company in FDA record
- GE OEC Medical Systems, Inc
- FDA record creation date
- Aug 23, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A firmware issue in the supplied LCD workstation monitor(s). Under some circumstances, the monitor(s) may appear dark with no video displayed.
Source facts
- Action
- On July 22, 2016 GE Healthcare Surgery began distributing an Urgent Medical Device Correction letter to customers via mail to inform them of the issue and provide instructions for immediate mitigation of the issue. The letter was sent to 3 titles at each facility affected: 1.) Facility Administrator, 2.) Director/Manager of Radiology and Radiologists, 3.) Radiology Department. Consignees instructured that local GE Healthcare Field Engineer will contact them to coordinate replacement of the affected monitor(s), and will ensure that system meets specifications.
- Initiated
- 2016-07-22
- Root cause
- Software design (manufacturing process)
- Terminated
- 2016-12-19
- Distribution
- Distribution US nationwide and Korea.
- Recall number
- Z-2628-2016
- Recall status
- Terminated
- CDRH record ID
- 148501
- FDA record created
- 2016-08-23
- Recall event number
- 74808