Skip to content

Public FDA record

Z-2628-2017: Merge LIS software. The firm name on the label is Merge Healthcare. Merge LIS system is a complete syst

Recall record Z-2628-2017 for Merge Healthcare, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2628-2017
Company in FDA record
Merge Healthcare, Inc.
FDA classification date
May 25, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There are potential issues with results reporting for certain run-based tests. Under certain conditions, the wrong results could inadvertently be verified.

Source facts

Action
Merge Healthcare issued a letter dated April 15, 2017, via email or via certified mail if they did not have an email address for the customer. The letter notifies the customer of the issue, informs them of a workaround, and lets them know a fix is being planned and they will be notified once it is made available. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope. The response form asks, in part, if they are interested in accepting the fix. If they decline the fix, they are asked to provide the reason why. The response was required no later than April 29, 2016. For further questions, call (877) 741-5369.
Initiated
2016-04-15
Root cause
Software design
Terminated
2019-03-13
Distribution
Nationwide Distribution to facilities and distributors. Distribution was made to the Virgin Islands. Foreign distribution was also made. There was no military/government distribution.
Recall number
Z-2628-2017
Recall status
Terminated
CDRH record ID
155647
FDA classified
2017-05-25
Recall event number
77299

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗