Z-2628-2023: Ambu aView 2 A Monitor-The Ambu aView 2 Advance is intended to display live imaging data from compatible

Recall record Z-2628-2023 for Ambu Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2628-2023
Company in FDA record
Ambu Inc.
FDA classification date
Sep 22, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Multiple impacts, for example from being dropped, can cause to the Ambu aView 2 Advance to short-circuit and under special circumstances make the lithium-ion battery in the device combust leading to smoke and flames

Source facts

Action
Ambu Inc. (US Distirbutor) issued Urgent Field Safety Notice Instruction for Use update for aView 2 Advance on July 21, 2023 delivered via sales representative. Letter states reason for recall, health risk and action to take: The information in this Field Safety Notice is therefore only related to the earliest configurations (versions) of Ambu aView" 2 Advance as listed above. Please communicate this information to relevant personnel within your organization. Included with the Field Safety Notice, you will find an insert for the Instruction for Use for Ambu aView" 2 Advance. The insert should be read and kept together with Instruction for Use you received together with your Ambu aView" 2 Advance. The information is also included in Appendix 2 of this notice. Ambu A/S is not removing any Ambu aView 2" Advance from the field; devices remain available for use. Within one month of receipt of this letter, please return confirmation of receipt of this Field Safety Notice (appendix 1). The traceability system at Ambu indicates that your institution has purchased an Ambu aView" 2 Advance within the affected configurations. You should check the version number and manufacturing date (as indicated above) of your device to identify if you have a potentially affected device within your facility. In the case that your device is related to the earliest configurations of Ambu aView" 2 Advance, you address this by familiarizing yourself with the information in the Instruction for Use insert and keep the insert together with the Instruction for Use. This notice needs to be passed on to all those who this might concern within your organization or to any organization where the devices could have been transferred. Please transfer this notice to other organizations on which this action has an impact. Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Patient safety remains our highest priority.
Initiated
2023-07-21
Root cause
Device Design
Distribution
Nationwide
Recall number
Z-2628-2023
Recall status
Open, Classified
CDRH record ID
202813
FDA classified
2023-09-22
Recall event number
92886

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗