Z-2629-2023: Aquilex Fluid Control System-Intended to provide fluid distension of the uterus during diagnostic and ope

Recall record Z-2629-2023 for WOM World Of Medicine AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2629-2023
Company in FDA record
WOM World Of Medicine AG
FDA classification date
Jan 5, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The display of inflow volume can reach its limit of 30,000 ml during long procdeure and the inflow volume display will freeze at the maximum value while the deficit will start counting backwards until 0 ml is reached and may result in fluid overload

Source facts

Action
W.O.M. World of Medicine GmbH (WOM), notified the US/Canada distributor Hologic via E-mail Urgent Field Safety Notice (FSN) - MEDICAL DEVICE RECALL Aquilex Fluid Control System Pump - REF AQL-100PBS and AQL-100P WOM reference no.: 2023-0001 on 7/17/23. Hologic notified sales representatives and the end users (hospitals and medical facilities) in the US and Canada on 8/21/2023 via Telephone calls and letter. Letter states reason for recall, health risk and action to take: W.O.M. WORLD OF MEDICINE GmbH has developed a software version, which, among others, starts emitting visual and acoustic signals when an inflow volume of 28,000 ml is reached. Additionally, the system will pause when an inflow volume of 30,000 ml is reached and is designed to prepare the user for a manual fluid deficit determination. Action To Be Taken by the User: Ensure that the affected devices receive their preventive maintenance on time, every two years, as prescribed by the manufacturer. A new software version and an updated version of the IFU are implemented during the next due preventive maintenance of the pump of the Aquilex Fluid Control Systems AQL-100PBS as well as AQL-100P. Also, please complete Customer Acknowledgement Form (Annex 2). This notice should be provided to all personnel who need to be made aware of this issue within your organization.
Initiated
2023-07-17
Root cause
Software design
Distribution
Nationwide Foreign: Canada, Australia, New Zealand, Belgium, France, Germany, Ireland, Netherlands, Spain, Sweden, Norway, Switzerland, UK, Kuwait, Qatar, Turkey, UAE
Recall number
Z-2629-2023
Recall status
Open, Classified
CDRH record ID
202951
FDA classified
2024-01-05
Recall event number
92938

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗