Z-2630-2011: 1500T12 Cool Path Ablation System - Generator The generator is a microprocessor-controlled device that p

Recall record Z-2630-2011 for St Jude Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2630-2011
Company in FDA record
St Jude Medical
FDA classification date
Jun 23, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The recall was initiated because St. Jude Medical has confirmed its T14 and T12 RF Generator may malfunction upon encountering an unexpected electric current (e.g. static electricity). The resulting malfunction results in a failure of the generator to respond to key presses properly, and may cause unexpected energy delivery or inability to discontinue energy delivery with the start / stop butt

Source facts

Action
The firm, St. Jude Medical (SJM), sent a "Product Correction" letter dated May 9, 2011 to it's affected customers. The letter described the product, problem and actions to be taken. The customers were instructed to discontinue use of the T14/T12 generators and remote controls until the units have been properly serviced; complete and return the attached field action form, and utilize the units in accordance with the Operators Manual. In case of system malfunction, the customers were instructed to shut off the RF power delivery by turning off the generator via the rocker switch, dislodging the power cord, or disconnecting of the patient cable or indifferent electrode connection. An SJM sales representative will contact the customers in the near future to facilitate repair or exchange of the impacted unit. If you have questions regarding this notice, contact the Marketing Manager, AF at (905-286-4015), Marketing Director, AF/CRM, at (905-286-4107); Product Manager (Tokyo) 03-6255-5740, and/or your SJM sales representative.
Initiated
2011-05-09
Root cause
Other
Terminated
2012-04-03
Distribution
International distribution only: Canada and Japan.
Recall number
Z-2630-2011
Recall status
Terminated
CDRH record ID
100432
FDA classified
2011-06-23
Recall event number
58859

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗