Z-2630-2016: 120 V Neptune 3 Rover, Model Number: 0703-001-000

Recall record Z-2630-2016 for Stryker Instruments Div. of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2630-2016
Company in FDA record
Stryker Instruments Div. of Stryker Corporation
FDA classification date
Aug 23, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Stryker initiated a voluntary recall of the 120 V Neptune 3 Rover on 07/01/2016, due to a potential for the device to experience system errors causing the device to shutdown if high levels of electromagnetic interference are present in the operating room.

Source facts

Action
Stryker Instruments initiated a voluntary recall of the Stryker Neptune 3 Rover via certified mail on 07/01/2016, due to the potential for the Neptune 3 Rover to experience system errors if there are high levels of electromagnetic interference in the operating room. 1. Immediately review the Recall Notification. 2. Confirm that the Rover serial numbers identified on the Business Reply Form (BRF) are at your facility. 3. A Stryker representative will facilitate the repair of your affected Neptune 3 Rover(s). Please coordinate the availability of the product to assure the repair can be implemented. 4. Please complete and return the Business Reply Form (BRF). Note: Your signature on the BRF indicates that you received and understand this Notification and have followed the instructions in the Notification. 5. Fax the completed, signed Business Reply Form to Stryker Instruments Regulatory Department, 866-521-2762, or scan and email a copy to kellyjo.davis@stryker.com. 6. Stryker will document the repair of each unit by serial number. For questions regarding this recall, please contact Stryker Instruments: Kelly Jo Davis 269-389-2921 kellyjo.davis@stryker.com
Initiated
2016-07-01
Root cause
Under Investigation by firm
Terminated
2017-06-02
Distribution
Domestic:MI, CA, ID, IA VA/DOD: None Foreign: None
Recall number
Z-2630-2016
Recall status
Terminated
CDRH record ID
148396
FDA classified
2016-08-23
Recall event number
74759

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗