Skip to content

Public FDA record

Z-2631-2017: Pointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range Standard Set 5

Recall record Z-2631-2017 for Medtest Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2631-2017
Company in FDA record
Medtest Holdings, Inc.
FDA classification date
Jun 26, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Individual vials of calibrator and standard were mislabeled; specifically the units on the vial labels are incorrectly listed as mg/dl when the correct unit is mg/L.

Source facts

Action
Technical Bulletin sent notification on July 21, 2009, to affected consignees via e-mail, fax and US Mail. The recall notification included a description of the reason for the recall, affected product, and consignee responsibilities. Customers with questions were instructed to contact Pointe Scientific's Technical Service Department at 1-800-757-5313. For questions regarding this recall call 800-757-5313.
Initiated
2009-07-21
Root cause
Error in labeling
Terminated
2017-07-24
Distribution
Worldwide Distribution - US including CA, MS, NJ, PA, HI, MD, FL, OK, NV. Foreign Distribution to CHILE
Recall number
Z-2631-2017
Recall status
Terminated
CDRH record ID
156076
FDA classified
2017-06-26
Recall event number
77451

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗