Z-2635-2023: Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004.

Recall record Z-2635-2023 for Fresenius Kabi USA, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2635-2023
Company in FDA record
Fresenius Kabi USA, LLC
FDA classification date
Sep 25, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set

Source facts

Action
Ivenix initiated contact on 4/22/22/ via phone communicating the contents of the customer letter which was then emailed to each customer. Letter states reason for recall, health risk and action to take: Please note, the specified start-up flow accuracy of up to +8% is maintained throughout the extended start-up time; and the steady-state flow accuracy of +l-SYo of the LVP is not affected with this condition. There is no corrective action you need to take at this time. We believe that this condition is low risk in that the affected flow rate range is limited, the start'up flow rate accuracy is unaffected and still within specifications, and the clinical practice of monitoring patients while administering high alert drugs mitigates the risk. Ivenix is working on a software update to correct this condition. It will be available in approximately 60 days. We will inform you when that update is ready to install. If you have additional questions, please contact your Ivenix representative or support below. Customer support - email Customer support - phone suplrort@ivenix.com I-978-775-8TAA
Initiated
2022-04-22
Root cause
Device Design
Distribution
US Nationwide distribution in the states of NJ, WI.
Recall number
Z-2635-2023
Recall status
Open, Classified
CDRH record ID
203046
FDA classified
2023-09-25
Recall event number
92973

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗