- FDA identifier
- Z-2635-2023
- Company in FDA record
- Fresenius Kabi USA, LLC
- FDA classification date
- Sep 25, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set
Source facts
- Action
- Ivenix initiated contact on 4/22/22/ via phone communicating the contents of the customer letter which was then emailed to each customer. Letter states reason for recall, health risk and action to take: Please note, the specified start-up flow accuracy of up to +8% is maintained throughout the extended start-up time; and the steady-state flow accuracy of +l-SYo of the LVP is not affected with this condition. There is no corrective action you need to take at this time. We believe that this condition is low risk in that the affected flow rate range is limited, the start'up flow rate accuracy is unaffected and still within specifications, and the clinical practice of monitoring patients while administering high alert drugs mitigates the risk. Ivenix is working on a software update to correct this condition. It will be available in approximately 60 days. We will inform you when that update is ready to install. If you have additional questions, please contact your Ivenix representative or support below. Customer support - email Customer support - phone suplrort@ivenix.com I-978-775-8TAA
- Initiated
- 2022-04-22
- Root cause
- Device Design
- Distribution
- US Nationwide distribution in the states of NJ, WI.
- Recall number
- Z-2635-2023
- Recall status
- Open, Classified
- CDRH record ID
- 203046
- FDA classified
- 2023-09-25
- Recall event number
- 92973