Open original FDA record ↗Recall record
- FDA identifier
- Z-2636-2025
- Company in FDA record
- Maquet Cardiovascular, LLC
- FDA classification date
- Sep 25, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.
Source facts
- Action
- Getinge Field Representatives contacted impacted accounts by telephone on 8/6/25 to notify them of this recall action and provide instructions for quarantine and return of affected devices. Consignees were asked to stop all visual inspection and return all HemoPro 3 units. Returns can be coordinated by calling or emailing Michele Kanis for an RMA at michele.kanis@getinge.com or (408)635-0647. Consignees are to utilize HemoPro 2 in place of HemoPro 3 units. Getinge Field Representatives delivered an updated URGENT MEDICAL DEVICE RECALL notice starting on 10/6/25 dated October 2025. The written notification is a supplement to the verbal notification made previously.
- Initiated
- 2025-08-06
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution in the states of FL, IL, MD & TX.
- Recall number
- Z-2636-2025
- Recall status
- Open, Classified
- CDRH record ID
- 215660
- FDA classified
- 2025-09-25
- Recall event number
- 97469