Siemens Medical Solutions USA, Inc - Radiation Oncology
FDA classification date
Jun 23, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
The firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis.
Source facts
Action
The firm, SIEMENS, sent a "Customer Information" letter to its customers on November 4, 2010. The letter describes the product, problem and actions to be taken. The customers were instructed to include this letter in their BEAMVIEW TI System Owner Manual chapter "Safety Advisory Letters" where it should remain. Siemens has developed an updated kit to install an additional grounding cable on the camera's central control unit to eliminate the problem. The updated kit was delivered with the Customer Information letter.
If you have any questions, contact the Service Manager- Regulatory Affairs at 925-246-8407.
Initiated
2010-11-04
Root cause
Other
Terminated
2011-11-14
Distribution
Worldwide distribution:
Recall number
Z-2637-2011
Recall status
Terminated
CDRH record ID
100351
FDA classified
2011-06-23
Recall event number
58813