Open original FDA record ↗Recall record
- FDA identifier
- Z-2637-2014
- Company in FDA record
- Data Innovations, LLC
- FDA classification date
- Sep 12, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential issues with the ProVue driver have been identified which may cause data to be misassociated with patient results under a given set of circumstances.
Source facts
- Action
- Data Innovations LLC sent an Important Medical Device Information letter on August 15, 2014, to consignees for the JResultNet using ProVue Instrument Driver JResultNet (provue.jrm) v. 1.5.3 and below. Potential issues with the ProVue driver have been identified which may cause data to be mis-associated with patient results under a given set of circumstances. The user on the ProVue transmits the same results twice back to back by clicking the Send to Host button twice. The ProVue sends two ASTM messages back to back before ending transmission (sending an EOT character) The solution was to disconnect JResultNet from the ProVue instrument and apply the patch to driver provue.jrm when it is available. If you have further questions regarding this notice, please feel free to contact us at northamerica-support@datainnovations.com or 239-931-6004 Toll-Free (from the US and Canada): 1-800-332-0499.
- Initiated
- 2014-08-15
- Root cause
- Software design
- Terminated
- 2015-03-04
- Distribution
- Nationwide Distribution including AL, AZ, CA, CT, DC, FL, GA, IN, LA, MI, MO, NC, NH, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, and WA.
- Recall number
- Z-2637-2014
- Recall status
- Terminated
- CDRH record ID
- 129603
- FDA classified
- 2014-09-12
- Recall event number
- 69086