Z-2637-2024: McGRATH" MAC EMS Video Laryngoscope, REF: 300-200-000 , Non-Sterile, Rx Only

Recall record Z-2637-2024 for Covidien. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2637-2024
Company in FDA record
Covidien
FDA classification date
Aug 21, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Battery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.

Source facts

Action
On July 17, 2024, Medtronic issued a "Urgent Medical Device Removal and Correction" notification to affected consignees via Mail and/or E-Mail. Medtronic ask consignees to take the following actions: Immediately discontinue use of the devices being removed (300-000-000 and 300-200-000). 1. Uninstall and dispose of the battery assembly from the devices being removed. 2. Return the removed device (300-000-000 and 300-200-000) to Medtronic. 3. Pass this notice on to all who need to be aware within your organization or to any organization where the potentially affected product with the specified serial numbers have been transferred or distributed. 4. Please complete the enclosed Customer Confirmation Form even if you do not have affected inventory. On 9/11/24, updated removal notices were mailed or emailed to customers expanding the recall to include affected video laryngoscopes shipped prior to 2019 that are now past their 5-year useful life and the batteries should be removed, devices should be removed from service and disposed of. The other actions above are the same.
Initiated
2024-07-09
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico, Virgin Islands and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Bhutan, Bosnia, Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Cameroon, Canada, Canary Islands, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican, Rep. Dutch Antilles, Egypt, El Salvador, Estonia, Faroe Islands, Fiji, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Islamic, Republic of Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania Luxembourg Macao Malaysia Maldives Malta Martinique Mauritius Mexico Moldova, Republic of Mongolia, Morocco, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion Romania, Russian Federation, Rwanda, Saint Martin, Saint Pierre And Miquelon, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Togo Trinidad And Tobago, Tunisia, Turkey, Turkmenistan, Uganda, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Viet Nam, Virgin Islands, U.S. Wallis And Futuna.
Recall number
Z-2637-2024
Recall status
Open, Classified
CDRH record ID
209174
FDA classified
2024-08-21
Recall event number
95033

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗