Skip to content

Public FDA record

Z-2637-2025: JUNCTIONAL EMERGENCY TREATMENT TOOL (JETT), REF: 30-0088, a component inside kits: KIT, TCCC Training-

Recall record Z-2637-2025 for North American Rescue LLC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2637-2025
Company in FDA record
North American Rescue LLC.
FDA classification date
Sep 26, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Device used for unilateral or bilateral occlusion of blood flow to the lower limbs has a potential manufacturing defect that may compromise the structural integrity of the device and device performance during use.

Source facts

Action
On 7/1/2025, recall notices were emailed to customers who were asked to do the following: 1) Quarantine the junctional emergency treatment tools. 2) Discontinue use and distribution. 3) Complete and return the response form via email to recalls@narescue.com In addition, if you have further distributed affected products, identify your customers and notify them of this recall. Your notification to your customers may be enhanced by including a copy of the recall notice. Instruct your customers to return affected product to you and then you will coordinate return with the recalling firm. This recall should be carried out to the user level. If you have questions contact the firm's customer service at 1-888-689-6277
Initiated
2025-07-01
Root cause
Nonconforming Material/Component
Distribution
Worldwide - US Nationwide distribution in the states of SC, OK, ID, KY, VA, NV, IL, TX, CA, FL, LA, NY, LA, ME, PA, AR, AZ, AL, DE, MA, MS, HI, WA, MI, GA, NJ, UT, MD, OR, NC, NC, AL and the countries of KOR, JPN, BWA, SAU, BEL.
Recall number
Z-2637-2025
Recall status
Open, Classified
CDRH record ID
215162
FDA classified
2025-09-26
Recall event number
97308

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗