Open original FDA record ↗Recall record
- FDA identifier
- Z-2638-2023
- Company in FDA record
- 3M Company - Health Care Business
- FDA classification date
- Sep 26, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132C and 135C.
Source facts
- Action
- 3M issued an Urgent product recall notice to its consignees on 08/24/2023 by email and US mail. The notice explained the problem with the product and requested the following: Discontinue distribution of the product and dispose of the product. Distributors were directed to notify their customers. The firm is offering replacement with the return of the completed response form.
- Initiated
- 2023-08-22
- Root cause
- Process control
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.
- Recall number
- Z-2638-2023
- Recall status
- Open, Classified
- CDRH record ID
- 202895
- FDA classified
- 2023-09-26
- Recall event number
- 92915