Z-2638-2026: Adapter for Head Positioning, Model/Catalog Number: 1739994, when used in conjunction with the side rails

Recall record Z-2638-2026 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2638-2026
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Jul 1, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Adapter for Head Positioning accessory may become loose during patient positioning and unintentionally move when used in conjunction with the side rails of the Upper Back Section H V U component on Baxter operating tables.

Source facts

Action
An URGENT PRODUCT RECALL notification letter dated 4/20/26 was sent to customers. Actions to be Taken by Customers 1. Please notify all users of the product affected within your facility, especially all operating room staff. The affected Adapters may continue to be used only when mounted directly to the side rails of the operating table back plate. Do not use the Adapter on the Upper Back Section H V U side rails (component). Using the Upper Back Section H V U (separate component which is attached to the operating table s back plate) in conjunction with the Adapter accessory must be stopped/ceased, until a replacement takes place. 2. Once the newly designed Adapters are available Baxter will contact customers with instructions for return and replacement of the impacted Adapters. 3. If you received this communication directly from Baxter, acknowledge receipt by following the within the body of the corresponding email. Acknowledging receipt of this notification will prevent you from receiving repeated notices. If you do not complete the acknowledgement, you will receive a phone call from Accenture LLP on behalf of Baxter to confirm your receipt of this notification. 4. If you purchased this product from a distributor or wholesaler, please contact your supplier to arrange for return of the affected product. Please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 5. Please forward a copy of this communication to the director of nursing, facility risk manager, patient safety, biomedical engineering department, director of surgical services, operating room staff, and any other departments within your institution who use the affected product. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please conduct a user-level recall of
Initiated
2026-04-21
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of Minnesota, Maryland, and Pennsylvania and the countries of Canada, Hong Kong, India, Japan, Korea, Philippines, Taiwan, Thailand, Vietnam, Austria, Belgium, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iraq, Israel, Italy, Jordan, Kenya, Lebanon, Libya, Malta, Morocco, Netherlands, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Saudi Arabia, Senegal, Serbia, Slovakia, Spain, Sweden, Switzerland, Tunisia, Turkey, United Arabia Emrites, Ukraine, Argentina, Brazil, Chile, Colombia, Costa Rica, Mexico, and Peru.
Recall number
Z-2638-2026
Recall status
Open, Classified
CDRH record ID
219911
FDA classified
2026-07-01
Recall event number
98833

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗