Z-2639-2024: C-Flow Bag Decanter, REF 2000S. Packaged on a form, fill and seal machine with a poly/Tyvek material then

Recall record Z-2639-2024 for Microtek Medical Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2639-2024
Company in FDA record
Microtek Medical Inc.
FDA classification date
Aug 16, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During sterile barrier testing performed on the decanter product line, the samples were identified to have pin holes and tears in the packaging film.

Source facts

Action
Consignees were notified via email and letter on 06/03/2024. The consignees were instructed to "Segregate Recalled Product/immediately examine the inventory for the lots listed above and remove all products from inventory regardless of its location and segregate this product in a secure location. The lot number can be found on the individual package or case label. If the consignee has affected product in their possession, please dispose of the product and packaging in accordance with hospital, administrative and /or local government policy. If the consignee is unable to dispose of the product, please contact the Customer Service team at the information provided below in this letter. The consignee is requested to please complete the enclosed Ecolab Customer Response Form and e-mail it to recall@ecolab.com, whether they have affected product in their possession or not. If they need assistance with filling out the form, please contact the sales representative. If the consignee has further distributed this product, instead complete the enclosed Ecolab Distributor Response Form and e-mail it to recall@ecolab.com whether their customers have affected product in their possession or not. Please identify the consignee's customers and notify recall@ecolab.com so Ecolab may notify them of this Field Recall. It is important that the firm receive the completed Ecolab Customer Response Form or Ecolab Distributor Response Form by June 14, 2024. The consignee is instructed to please contact the Customer Service at 1-800-824-3027 from 8:30 a.m. to 5 p.m. (CST) to obtain a Returned Material Authorization (RMA) number if they are unable to dispose of the affected product, or email at csdomesticorders@ecolab.com or cs-distributorrequests@ecolab.com. The consignee is to package the decanter(s) that are being returned in an appropriate shipping box. Place a copy of the response form in the box, appropriately seal the box, and send it on or before June 14, 2024, to: ATTN: Belissa del Val
Initiated
2024-06-03
Root cause
Device Design
Distribution
Distributed nationwide and internationally to Canada, Mexico, New Zealand, Belize, and Hong Kong.
Recall number
Z-2639-2024
Recall status
Open, Classified
CDRH record ID
208261
FDA classified
2024-08-16
Recall event number
94780

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗