Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.
Source facts
Action
On 05/23/2020, Customer Notification Letters dated 5/22/2020 were emailed to customers. Customers were asked to do the following:
For safety issue 1:
If you observe image artifacts, please reboot the system and continue the reconstruction. If the image artifacts persist, please contact the recalling firm's service support team or call the Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct).
For safety issue 2:
If the scout scanning is interrupted, try repeating the scout scanning. If problem persists, please contact the recalling firm's service support team or call the Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct).
Customers were asked to notify all users of the affected products via Customer Notification Letter.
On 8/30/2024, a correction notice was sent identifying an additional serial number.
Initiated
2020-05-23
Root cause
Software design
Distribution
US Distribution to state of: TX
Recall number
Z-2641-2020
Recall status
Open, Classified
CDRH record ID
181751
FDA classified
2020-07-20
Recall event number
85771