Z-2641-2020: Computed Tomography X-ray System, Model: uCT 550, GTIN: 06971576831036 Intended use The uCT Computed To

Recall record Z-2641-2020 for Shanghai United Imaging Healthcare Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2641-2020
Company in FDA record
Shanghai United Imaging Healthcare Co., Ltd.
FDA classification date
Jul 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.

Source facts

Action
On 05/23/2020, Customer Notification Letters dated 5/22/2020 were emailed to customers. Customers were asked to do the following: For safety issue 1: If you observe image artifacts, please reboot the system and continue the reconstruction. If the image artifacts persist, please contact the recalling firm's service support team or call the Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct). For safety issue 2: If the scout scanning is interrupted, try repeating the scout scanning. If problem persists, please contact the recalling firm's service support team or call the Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct). Customers were asked to notify all users of the affected products via Customer Notification Letter. On 8/30/2024, a correction notice was sent identifying an additional serial number.
Initiated
2020-05-23
Root cause
Software design
Distribution
US Distribution to state of: TX
Recall number
Z-2641-2020
Recall status
Open, Classified
CDRH record ID
181751
FDA classified
2020-07-20
Recall event number
85771

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗