Open original FDA record ↗Recall record
- FDA identifier
- Z-2642-2026
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jul 1, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
Source facts
- Action
- Siemens notified consignees on about 05/19/2026 via letter. Consignees were instructed to ensure all users of the affected products within each organization and others who may need to be informed will receive the relevant safety information provided within the notice and comply with the recommendations therein, maintain the recall notification with product related records and until the corrective measures have been completed. Siemens will be releasing a software update to correct the issue via an Update Instruction AX018/26/S. All affected systems will be corrected by Siemens Healthineers. The service organization will contact consignees in the near future to arrange a date to perform the corrective action. Additionally, consignees were instructed to confirm and acknowledge receipt of the recall notification digitally.
- Initiated
- 2026-05-20
- Root cause
- Software design
- Distribution
- US Nationwide distribution.
- Recall number
- Z-2642-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220754
- FDA classified
- 2026-07-01
- Recall event number
- 99079