Open original FDA record ↗Recall record
- FDA identifier
- Z-2644-2016
- Company in FDA record
- Ziehm Imaging Inc
- FDA record creation date
- Aug 25, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.
Source facts
- Action
- Ziehm Imaging, Inc planned action to bring these defects into compliance: 1. You will contact customers and initiate a software update to correct the defects. 2. You will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. For further questions, please call (866) 949-4346.
- Initiated
- 2016-04-15
- Root cause
- Device Design
- Terminated
- 2019-07-18
- Distribution
- US Distribution including Puerto Rico and to the states of :FL, GA, VA,MA. ME, MI, NY, NJ, PA, RI, IL, TX, AZ, CA, NV, UT, and WA.
- Recall number
- Z-2644-2016
- Recall status
- Terminated
- CDRH record ID
- 145918
- FDA record created
- 2016-08-25
- Recall event number
- 74013