Open original FDA record ↗Recall record
- FDA identifier
- Z-2644-2017
- Company in FDA record
- Exactech, Inc.
- FDA classification date
- Jun 19, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to perform calcar planing to the bone, i.e., to include a caution statement about the potential for instrument breakage if bending or torsional forces are applied.
Source facts
- Action
- Exactech sent an Field Advisory Notice dated May 22, 2017, by mail informing the customers the updated surgical technique is available on Exactech's website (www.exac.com). The notice requested consignees: Extend the information to accounts in possession of the product Complete & return the attached Field Advisory Notice Response Form to Exactech. For further questions, please call (800) 392-2832.
- Initiated
- 2017-05-22
- Root cause
- Under Investigation by firm
- Terminated
- 2019-07-26
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AL, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, RI, SC, TN, TX, VA, WA, WI, Hawaii., and to the countries of : Argentina, Australia, Austria, Brazil, China, Colombia, France, Germany, Guatemala, India, Italy, Japan, Spain, Switzerland, Turkey, and United Kingdom
- Recall number
- Z-2644-2017
- Recall status
- Terminated
- CDRH record ID
- 154804
- FDA classified
- 2017-06-19
- Recall event number
- 77002