Z-2646-2023: BD Pyxis MedBank MedPass Software, REF: 139088-01

Recall record Z-2646-2023 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2646-2023
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Sep 28, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

After dispensing a timed dose medication for a patient, the automated dispensing cabinet software may continue to show the already dispensed medication dose is still available for dispense, which could lead to duplicate medication administration.

Source facts

Action
On 8/18/23, correction notices were mailed to directors of Biomedical Engineering/Risk Management/Nursing/Pharmacy and distributors. They were asked to take the following actions: 1) Remind users that affected devices do not replace the electronic (or paper) patient record, specifically the medication order, which will remain accurate and serves as the source of truth for medication administration. Nursing best practice is to confirm the 5 rights of medication orders prior to administration. Some inpatient psychiatric units also utilize barcode medication administration (BCMA) which may detect duplicate administrations. If a duplicate medication is inadvertently administered, ensure the patient is appropriately monitored and follow organizational process for reporting of a medication error. 2. Please ensure this letter is forwarded to applicable facilities if devices have been transferred and that all appropriate personnel have been made aware of this issue. 3. Complete and return the response form. Adverse health consequences experienced with the use of this product should be reported to BD at 1-800-727-6102 or via bd.com/self-service The recall issue will be resolved in software v1.2, which will be provided to all customers approximately October 2023 via remote upgrade. On 8/26/24, correction-update notices were sent informing customers the field correction (software update to BD Pyxis MedBank MedPass software v1.2.1.17) was not effective. As a result, devices required an additional update to BD Pyxis MedBank MedPass software version v1.2.3.23.
Initiated
2023-08-18
Root cause
Device Design
Distribution
US: FL, MI, KS
Recall number
Z-2646-2023
Recall status
Open, Classified
CDRH record ID
203032
FDA classified
2023-09-28
Recall event number
92967

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗