- FDA identifier
- Z-2647-2023
- Company in FDA record
- Baxter Healthcare Corporation
- FDA classification date
- Sep 28, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Epiphany, now a Baxter Healthcare company, is issuing an Urgent Medical Device Correction for the Cardio Server E3 ECG Management System software listed below due to a report of the caliper tool becoming non-responsive, preventing the caliper function from being used temporarily on software versions v6.1.x, v6.2.x, and v7.0.x of the E3 user interface.
Source facts
- Action
- Baxter Healthcare initially notified consignees via email on 08/10/2023 of the issue and that a software patch would be implemented. An additional notification was sent to consignees via first-class mail on 08/30/2023. Consignees were instructed that Cardio Server support representatives will contact each facility to arrange for the installation of the software patch or consignees can proactively reach out to the Cardio Server Support team to schedule the patch update. Operators may continue to use the Cardio Server ECG Management System caliper tool prior to the installation of the software patch by following the steps listed in the notification letter. Customers were also instructed to notify other facilities or departments if the affected units have been further distributed, and to acknowledge receipt of the notification by responding on Baxter's portal at the website included in the letter.
- Initiated
- 2023-08-10
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AL, AR, AZ, CA, CA, CT, FL, FL, GA, IA, ID, IL, IN, KS, KY, MA, MD, MO, MS, MT, NE, NH, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Canada and Bermuda.
- Recall number
- Z-2647-2023
- Recall status
- Open, Classified
- CDRH record ID
- 202953
- FDA classified
- 2023-09-28
- Recall event number
- 92940