Public FDA record

Z-2647-2023: Baxter Epiphany Cardio Server E3 ECG Management System Servers with software versions: a) v6.1.x, b) v6.

Recall record Z-2647-2023 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2647-2023
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Sep 28, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Epiphany, now a Baxter Healthcare company, is issuing an Urgent Medical Device Correction for the Cardio Server E3 ECG Management System software listed below due to a report of the caliper tool becoming non-responsive, preventing the caliper function from being used temporarily on software versions v6.1.x, v6.2.x, and v7.0.x of the E3 user interface.

Source facts

Action
Baxter Healthcare initially notified consignees via email on 08/10/2023 of the issue and that a software patch would be implemented. An additional notification was sent to consignees via first-class mail on 08/30/2023. Consignees were instructed that Cardio Server support representatives will contact each facility to arrange for the installation of the software patch or consignees can proactively reach out to the Cardio Server Support team to schedule the patch update. Operators may continue to use the Cardio Server ECG Management System caliper tool prior to the installation of the software patch by following the steps listed in the notification letter. Customers were also instructed to notify other facilities or departments if the affected units have been further distributed, and to acknowledge receipt of the notification by responding on Baxter's portal at the website included in the letter.
Initiated
2023-08-10
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AL, AR, AZ, CA, CA, CT, FL, FL, GA, IA, ID, IL, IN, KS, KY, MA, MD, MO, MS, MT, NE, NH, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Canada and Bermuda.
Recall number
Z-2647-2023
Recall status
Open, Classified
CDRH record ID
202953
FDA classified
2023-09-28
Recall event number
92940

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗