Open original FDA record ↗Recall record
- FDA identifier
- Z-2647-2026
- Company in FDA record
- RAYSEARCH LABORATORIES AB
- FDA classification date
- Jul 1, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.
Source facts
- Action
- A Field Safety Notice, Medical Device Correction #175428 dated 6/2/26 was emailed to consignees. The notice advises consignees to ensure clear verbal or procedural communication when updating the beam set delivery note, avoid making changes from any location other than the control room when treatment is ongoing, avoid relying on beam set delivery note for conveying critical information. Additionally, consignees are to ensure planning staff and users are educated of the provided workaround. Consignees are to verify the software version of their units and reply to the emailed notification. A software update is scheduled for August 2026 to address this issue. Consignees with questions can contact quality@raysearchlabs.com.
- Initiated
- 2026-05-20
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide - US Nationwide distribution.
- Recall number
- Z-2647-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221026
- FDA classified
- 2026-07-01
- Recall event number
- 99141