Z-2648-2026: ARTIS icono ceiling. Model Number: 11328100.

Recall record Z-2648-2026 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2648-2026
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Jul 1, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor attached to the suspension may gradually loosen.

Source facts

Action
Siemens Healthineers notified consignees on about 06/08/2026 via "CUSTOMER SAFETY ADVISORY NOTICE" letter. Consignees were instructed to conduct regular inspections of the gap between the monitor carrier and the ceiling-mounted arm and be attentive to any unusual noises, shaking, wobbling, or other unexpected behavior during system operation, as these may indicate loose screws. If any such signs are observed, limit movements of the DCS and contact Siemens service organization immediately for further instructions. Always ensure that the DCS is positioned for safe use. All affected systems will be corrected by Siemens Healthineers service engineers. Siemens engineers will correct the screwed connection of the monitor ceiling support according to the adapted specifications and instructions. The correction will be released via an Update Instruction AX022/26/S. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented.
Initiated
2026-06-08
Root cause
Device Design
Distribution
US Nationwide distribution in the states of CA, GA, IA, IN, LA, MO, NC, NY, PA, SC.
Recall number
Z-2648-2026
Recall status
Open, Classified
CDRH record ID
221088
FDA classified
2026-07-01
Recall event number
99224

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗