Z-2649-2023: Centurion Sterile CirClamp with 1.45cm Bell Reprocessed, Reorder CR3145K, 1 per package, 12 per case.

Recall record Z-2649-2023 for MEDLINE INDUSTRIES, LP - Northfield. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2649-2023
Company in FDA record
MEDLINE INDUSTRIES, LP - Northfield
FDA classification date
Sep 29, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The kits incorrectly contain CirClamp subassembly 345CRSAK which results in the kits containing an incorrect bell and base size.

Source facts

Action
The recalling firm issued letters dated 5/23/2023 via email and first class mail on 5/23/2023 to its consignees, which identified incorrect information regarding the affected product. It had stated the label indicates the kit contains a 1.3 cm bell and insert; however the kit actually contains a 1.45 cm bell and insert. The affected kits were to be destroyed for credit and none were to be returned to the recalling firm. The consignee was to check their stock for the affected item number and lot numbers listed on the recalling firm's recall portal and destroy the affected product. A link was provided to complete the response form. If the consignee is a distributor or has resold or transferred the product to another company or individual, they are to notify them of this recall communication. The recalling firm issued a letter flagged as an Amended Letter dated 9/1/2023 via email and first class mail that alerted the consignee to the incorrect product information contained in the initial 5/23/2023 letter and the Amended Letter provided the correct product information, that being the product is supposed to include a 1.45 cm base and 1.45cm bell, however, the kit incorrectly contains CirClamp subassembly 345CRSAK which includes a 1.6cm base and 1.6cm bell. The letter instructs that affected kits can be destroyed for credit and no kits are to be returned to Medline. The consignee is instructed to immediately check their stock for the item number and lot number listed in the letter and on the destruction form, quarantine the affected product, and destroy it. The enclosed destruction form listing the quantity of affected product destroyed is to be returned. If the product has been transferred to another individual, department or location , the consignee is to notify them of this recall communication.
Initiated
2023-05-23
Root cause
Process control
Distribution
US Nationwide distribution.
Recall number
Z-2649-2023
Recall status
Open, Classified
CDRH record ID
202810
FDA classified
2023-09-29
Recall event number
92883

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗