Open original FDA record ↗Recall record
- FDA identifier
- Z-2651-2014
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Sep 16, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Beckman Coulter is recalling Coulter DxH Diluent because it may be associated with elevated platelet (PLT) background counts between 3-7 x 10(3) cells/uL. This will result in failed Daily Checks on the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems.
Source facts
- Action
- Beckman Coulter sent an Urgent Medical Device Recall letter dated August 25, 2014, to all customers that purchased the Coulter DxH Diluent. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to complete and return the enclosed response form within 10 days. Customers with questions were instructed to contact Beckman Coulter's Customer Support Center at http://www.beckmancoulter.com, or call (800) 526-7694 in the US and Canada.
- Initiated
- 2014-08-25
- Root cause
- Under Investigation by firm
- Terminated
- 2015-02-25
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Canada, Chile, Mexico, Panama, and Thailand. .
- Recall number
- Z-2651-2014
- Recall status
- Terminated
- CDRH record ID
- 129659
- FDA classified
- 2014-09-16
- Recall event number
- 69112