Open original FDA record ↗Recall record
- FDA identifier
- Z-2651-2023
- Company in FDA record
- Covidien
- FDA classification date
- Sep 29, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from being charged, and could lead to treatment delay.
Source facts
- Action
- On 8/22/23 Medtronic mailed recall notices to risk managers and directors of respiratory care informing them of the following: 1) Discontinue use, quarantine, and return affected devices to the firm. 2) Complete and return the customer confirmation form via email to it to rs.gmbfcamitg@medtronic.com. 3) Forward the notice to all those who need to be aware within your organization or to any organization where the potentially affected product with the specified serial numbers has been transferred or distributed. If you have any questions regarding this communication, contact customer service at 800-962-9888, Option 2. Quality problems can be reported to Quality Assurance at 800-255-6774 option 4.
- Initiated
- 2023-08-22
- Root cause
- Component design/selection
- Distribution
- US: AL, NJ, MA, MI, NY, NC, CO, TX, FL, KY, UT, CA, AZ, KS, IL, TN. OUS: El Salvador, Sweden, Canary Islands
- Recall number
- Z-2651-2023
- Recall status
- Open, Classified
- CDRH record ID
- 202954
- FDA classified
- 2023-09-29
- Recall event number
- 92941