Skip to content

Public FDA record

Z-2651-2023: Puritan Bennett Cuff Pressure Manager, REF: 180-03

Recall record Z-2651-2023 for Covidien. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2651-2023
Company in FDA record
Covidien
FDA classification date
Sep 29, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from being charged, and could lead to treatment delay.

Source facts

Action
On 8/22/23 Medtronic mailed recall notices to risk managers and directors of respiratory care informing them of the following: 1) Discontinue use, quarantine, and return affected devices to the firm. 2) Complete and return the customer confirmation form via email to it to rs.gmbfcamitg@medtronic.com. 3) Forward the notice to all those who need to be aware within your organization or to any organization where the potentially affected product with the specified serial numbers has been transferred or distributed. If you have any questions regarding this communication, contact customer service at 800-962-9888, Option 2. Quality problems can be reported to Quality Assurance at 800-255-6774 option 4.
Initiated
2023-08-22
Root cause
Component design/selection
Distribution
US: AL, NJ, MA, MI, NY, NC, CO, TX, FL, KY, UT, CA, AZ, KS, IL, TN. OUS: El Salvador, Sweden, Canary Islands
Recall number
Z-2651-2023
Recall status
Open, Classified
CDRH record ID
202954
FDA classified
2023-09-29
Recall event number
92941

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗