A customer experienced when using IMPAX CV Reporting software, specifically, when building a NIV report, the NIV Cardio report was showing incorrect findings for Aneurysm.
Source facts
Action
AGFA Healthcare sent an Urgent Field Safety Notice dated December 22, 2016, to all affected consignees. The notice included an acknowledgement fax-back or email indicating that the information was received and understood. The notice informed the consignees of the problem and indicated that they would be contacted by Agfa Service to check the problem and to correct the NIV reporting clinical content. Customers with questions were instructed to call 1-401-330-7433.
For questions regarding this recall call 864-421-1984.
Initiated
2016-12-22
Root cause
Software design
Terminated
2021-08-27
Distribution
Nationwide Distribution to NJ, NC, OH, PA, SC, TN, TX, and WI
Recall number
Z-2652-2017
Recall status
Terminated
CDRH record ID
156955
FDA record created
2017-06-29
Recall event number
77629