Z-2652-2023: Venue Go Standard Carts Ref: (H45181VC and H45103VCW), used with Venue Go R2, R3, R4 systems R2, R3, and

Recall record Z-2652-2023 for GE MEDICAL SYSTEMS, ISRAEL LTD.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2652-2023
Company in FDA record
GE MEDICAL SYSTEMS, ISRAEL LTD.
FDA classification date
Sep 29, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some Venue Go Standard Carts can develop an internal failure of the vertical/horizontal tilt adjustment mechanism which can result in the Venue Go system detaching from the cart and falling.

Source facts

Action
The recalling firm issued letters dated 9/6/2023 via FedEx in the U.S. which explained the safety issue and actions to be taken by the customer/user. The actions included: On a weekly basis, check the vertical/horizontal tilt adjustment mechanism on the Venue Go Standard Cart to ensure it is secure: (1) Lock the wheels; (2) Raise the up/down adjustment column to its highest position; (3) Observe the tilt mechanism during side-to-side rocking as is shown in the photos in the letter. If the vertical/horizontal tilt adjustment mechanism is loose, the consignee is to (1) remove the Venue Go system from the Cart cradle and place on a tabletop using the adjustable rear support stand; and (2) Do not use the Venue Go Standard Cart until the corrective action is performed by GE HealthCare. All potential users i the facility are to be made aware of the safety notification and recommended actions. An acknowledgement form was enclosed for completion and return to the firm. In addition, the letter provided specific affected product details along with photographs on how to identify the manufacture date on the Venue Go Standard Cart since the affected devices that were impacted were manufactured 2022-08 and before. GE HealthCare will correct all affected products and a representative will contact the consignee to arrange for correction.
Initiated
2023-09-06
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide distribution including government and military distribution as well as PR.
Recall number
Z-2652-2023
Recall status
Open, Classified
CDRH record ID
203365
FDA classified
2023-09-29
Recall event number
93069

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗